1. Summary

This adaptor is for medical use with input power ranging at 90-264Vac,single current output 24Vdc/2.5A, and peak current 6A. It provides overcurrent protection, short circuit protection, overvoltage protection and overheat protection.

 

 

 

2. Product’s Key Features

Model Input Voltage Range Rated Output Voltage Output Current Range

Ripple(Peak)

(Rated Load,bandwidth Limitation 20MHz)

MDA60-220S24 90—264Vac 24Vdc 0—2.5A ≤150mV

 

 

3. Adoption Standards and Guidance

  1. GB/T 2423.1—2008 Environmental Testing for Electrical and Electronic Products – Part 2: Test Methods/Test A: Low Temperature
  2. GB/T 2423.2—2008 Environmental Testing for Electrical and Electronics Products – Part 2: Test Methods/Test B: High Temperature
  3. GB/T 2423.3—2006 Basic Environmental Testing procedures for Electric and Electronic Products -- Test Ca: Damp Heat, Steady State
  4. GB/T 2423.4—2008 Basic Environmental Testing procedures for Electric and Electronic Products -- Test Db: Damp Heat, Cyclical
  5. GB/T 2423.5—2008 Environmental Testing for Electrical and Electronic Products – Part 2: Test Methods/Test Ea and Guidance: Shock
  6. GB/T 2423.6—1995 Environmental Testing for Electrical and Electronic Products – Part 2: Test Methods/Test Ea and Guidance:Collision
  7. GB/T 2423.8—2008 Environmental Testing for Electrical and Electronic Products – Part 2: Test Methods/Test Ed:Free Fall
  8. GB/T 2423.10—2008 Environmental Testing for Electrical and Electronic Products – Part 2: Test Methods/Test Fc and Guidance:Vibration (Sine)
  9. GB/T 14714.1—2008 Micro-computer Systems and Equipment General Specification of Switching Power Supply
  10. GB/T 17625.1—2012 Limits for harmonic current of low voltage electrical and electronic equipment.
  11. GB/T 17625.2-2007 Electromagnetic compatibility limits on voltage fluctuations and flicker created by devices with rated current less than 16A under low-voltage power supply systems.
  12. GB/T 17626.1-2006 Electromagnetic compatibility testing and measurement techniques overview of immunity test.
  13. GB/T 17626.11-2011 Electromagnetic compatibility - Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity test
  14. GB/T 17626.2-2006 Electromagnetic compatibility - Testing and measurement techniques - Electrostatic discharge immunity test
  15. GB/T 17626.3-2006 Electromagnetic compatibility - Testing and measurement techniques - RF electromagnetic field immunity test
  16. GB/T 17626.4-2008 Electromagnetic compatibility - Testing and measurement techniques - Electrical fast transient/burst immunity test
  17. GB/T 17626.5-2008 Electromagnetic compatibility - Testing and measurement techniques - Surge (impact) immunity test
  18. GB/T 17626.6-2008 Electromagnetic compatibility - Testing and measurement techniques – Immunity test for RF induction transmission to disturbance
  19. GB/T 17626.8-2006 Electromagnetic compatibility - Testing and measurement techniques - Power frequency magnetic field immunity test
  20. GB/T 17626.9-2011 Electromagnetic compatibility - Testing and measurement techniques - Pulse magnetic field immunity test
  21. GB/T 1772-1979 Test method for determination of failure rate of electronic elements and components
  22. QJ/T 1950-1990 Static operating system technical requirements
  23. GB/T 6991-1986 Representation of reliability data for electronic components
  24. MIL-HDBK-217F Electronic equipment reliability prediction
  25. GJB/Z-299B Electronic equipment reliability prediction manual
  26. GB9706.1-2007 Medical electronic equipment safety

 

4. Environmental Conditions

Serial No. Item Technical Indictor Unit Remarks
1 Working temperature 0 — +40  
2 Storing temperature -20 — +85  
3 Relative humidity Working 20—90 %RH  
Storing 10—95
4 Altitude ≤3000 m  
5 Cooling-down method Natural cooling    

 

5. Power Features

1. Input Features
Serial No. Item Technical Requirements Unit Remarks
1.1 Power type Category    
1.2 Rated input voltage 100—240 Vac  
Input voltage range 90—264
1.3 Input frequency range 47—63 Hz  
1.4 Rated input frequency 50/60 Hz  
1.5 Max input current 2 A 90VAC
1.6 Input inrush current ≤60 A VIN=240Vac input/ 25/cold on the first boot (Capacity should be released completely. Measure at any phase angle of input voltage.)
1.7 Power on delay time ≤2 S 220VAC/full load
1.8 Shutdown time ≥10 mS 110VAC/full load
1.9 No-load power consumption 2 W  
2. Output features
2.1 Output voltage range 24.00—24.60 Vdc Full input voltage, full load range
2.2
bigPhoto
 Mr. Sunny Cheng

Tel: 86-755-29724966
Contact to this supplier

Related Search

Find more related products in following catalogs on Hisupplier.com

Company Info

China Computer Peripherals Online Market [China (Mainland)]


Business Type:Manufacturer
City: Shenzhen
Province/State: Guangdong
Country/Region: China (Mainland)

You May Like:

Product (126)